1. Home
  2. OCTAID Home
  3. OCTAID Image Interpretation
  4. Post Stent Display Function

The stent display function is a recently developed tool which color codes malapposed stent struts based on the severity of the malapposed stent strut as seen in Figure A and Figure B. As indicated by the white box noted in Figure A, white, yellow and red colors are provided to each stent strut based on the degree of the malapposition. As depicted in the above in figures via the white arrows, these color coded regions appear on the angiogram and on the longitudinal and cross sectional OCT images. Simultaneous measurements can be made as well.

Ask the Experts

Take Quiz

Meet the Team

Contact Us

Case ExamplesEducational Content

ATHERECTOMY COMPLICATIONS

Case ExamplesEducational Content

BYPASS GRAFT COMPLICATIONS

ModelBurst (ATM)Diameter (mm)Balloon Length (mm)
681215202530
MINI TREK (CTO INDICATED)1.271.2✓✓✓✓✓
1.561.5✓✓✓✓✓
2.172✓✓✓✓✓✓*✓*
TREK2.372.25✓✓✓✓✓✓✓
2.672.5✓✓✓✓✓✓✓
2.912.75✓✓✓✓✓✓✓
3.113✓✓✓✓✓✓✓
3.53.25✓✓✓✓✓✓✓
3.73.5✓✓✓✓✓✓✓
3.943.75✓✓✓✓✓✓✓
4.324✓✓✓✓✓✓✓
4.864.5✓✓
5.365✓✓
*Only available as RX, not as OTW. Other sizes available in both versions.
Retrieved on April 07 2022 from Abbott product ordering information. Please look at the company's files for the latest available data on device configurations and avilability in your area.